Current closed-loop insulin delivery methods stem from sophisticated models of the glucose-insulin (G/I) system, mostly based on complex studies employing glucose tracer technology. We tested the performance of a new minimal model (GLUKINSLOOP 2.0) of the G/I system to characterize the glucose and insulin dynamics during multiple mixed meal tests (MMT) of different sizes in patients with type 1 diabetes (T1D) on insulin pump therapy (continuous subcutaneous insulin infusion, CSII). The GLUKINSLOOP 2.0 identified the G/I system, provided a close fit of the G/I time-courses and showed acceptable reproducibility of the G/I system parameters in repeated studies of identical and double-sized MMTs. This model can provide a fairly good and reproducible description of the G/I system in T1D patients on CSII, and it may be applied to create a bank of "virtual" patients. Our results might be relevant at improving the architecture of upcoming closed-loop CSII systems.

A Novel Insulin/Glucose Model after a Mixed-Meal Test in Patients with Type 1 Diabetes on Insulin Pump Therapy / Marchetti, Luca; Reali, Federico; Dauriz, Marco; Brangani, Corinna; Boselli, Linda; Ceradini, Giulia; Bonora, Enzo; Bonadonna, Riccardo C; Priami, Corrado. - In: SCIENTIFIC REPORTS. - ISSN 2045-2322. - ELETTRONICO. - 6:(2016), p. 36029. [10.1038/srep36029]

A Novel Insulin/Glucose Model after a Mixed-Meal Test in Patients with Type 1 Diabetes on Insulin Pump Therapy

Marchetti, Luca;Reali, Federico;Priami, Corrado
2016-01-01

Abstract

Current closed-loop insulin delivery methods stem from sophisticated models of the glucose-insulin (G/I) system, mostly based on complex studies employing glucose tracer technology. We tested the performance of a new minimal model (GLUKINSLOOP 2.0) of the G/I system to characterize the glucose and insulin dynamics during multiple mixed meal tests (MMT) of different sizes in patients with type 1 diabetes (T1D) on insulin pump therapy (continuous subcutaneous insulin infusion, CSII). The GLUKINSLOOP 2.0 identified the G/I system, provided a close fit of the G/I time-courses and showed acceptable reproducibility of the G/I system parameters in repeated studies of identical and double-sized MMTs. This model can provide a fairly good and reproducible description of the G/I system in T1D patients on CSII, and it may be applied to create a bank of "virtual" patients. Our results might be relevant at improving the architecture of upcoming closed-loop CSII systems.
2016
Marchetti, Luca; Reali, Federico; Dauriz, Marco; Brangani, Corinna; Boselli, Linda; Ceradini, Giulia; Bonora, Enzo; Bonadonna, Riccardo C; Priami, Cor...espandi
A Novel Insulin/Glucose Model after a Mixed-Meal Test in Patients with Type 1 Diabetes on Insulin Pump Therapy / Marchetti, Luca; Reali, Federico; Dauriz, Marco; Brangani, Corinna; Boselli, Linda; Ceradini, Giulia; Bonora, Enzo; Bonadonna, Riccardo C; Priami, Corrado. - In: SCIENTIFIC REPORTS. - ISSN 2045-2322. - ELETTRONICO. - 6:(2016), p. 36029. [10.1038/srep36029]
File in questo prodotto:
File Dimensione Formato  
srep36029.pdf

accesso aperto

Tipologia: Versione editoriale (Publisher’s layout)
Licenza: Creative commons
Dimensione 973.77 kB
Formato Adobe PDF
973.77 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11572/156167
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 8
  • OpenAlex ND
social impact