The goal of this study was to assess the imaging performances of the pCT system developed in the framework of INFN-funded (Italian National Institute of Nuclear Physics) research projects. The spatial resolution, noise power spectrum and RSP accuracy has been investigated, as a preliminary step to implement a new cross-calibration method for x-ray CT (xCT). Approach: The INFN pCT apparatus, made of four planes of silicon micro-strip detectors and a YAG:Ce scintillating calorimeter, reconstructs 3D RSP maps by a filtered-back projection algorithm. The imaging performances (i.e. spatial resolution, noise power spectrum and RSP accuracy) of the pCT system were assessed on a custom-made phantom, made of plastic materials with different densities ([0.66, 2.18] g/cm3). For comparison, the same phantom was acquired with a clinical xCT system. Main results: The spatial resolution analysis revealed the non-linearity of the imaging system, showing different imaging responses in air or water phantom background. Applying the Hann filter in the pCT reconstruction, it was possible to investigate the imaging potential of the system. Matching the spatial resolution value of the xCT (0.54 lp/mm) and acquiring both with the same dose level (11.6 mGy), the pCT appeared to be less noisy than xCT, with an RSP standard deviation of 0.0063. Concerning the RSP accuracy, the measured Mean Absolute Percentage Errors were (0.23+-0.09)% in air and (0.21+-0.07)% in water. Significance: The obtained performances confirm that the INFN pCT system provides a very accurate RSP estimation, appearing to be a feasible clinical tool for verification and correction of xCT calibration in proton treatment planning.
Characterization of the INFN proton CT scanner for cross-calibration of x-ray CT / Fogazzi, Elena; Trevisan, Diego; Farace, Paolo; Righetto, Roberto; Rit, Simon; Scaringella, Monica; Bruzzi, Mara; Tommasino, Francesco; Civinini, Carlo. - In: PHYSICS IN MEDICINE AND BIOLOGY. - ISSN 0031-9155. - 68:12(2023), p. 124001. [10.1088/1361-6560/acd6d3]
Characterization of the INFN proton CT scanner for cross-calibration of x-ray CT
Fogazzi, Elena;Trevisan, Diego;Tommasino, Francesco;
2023-01-01
Abstract
The goal of this study was to assess the imaging performances of the pCT system developed in the framework of INFN-funded (Italian National Institute of Nuclear Physics) research projects. The spatial resolution, noise power spectrum and RSP accuracy has been investigated, as a preliminary step to implement a new cross-calibration method for x-ray CT (xCT). Approach: The INFN pCT apparatus, made of four planes of silicon micro-strip detectors and a YAG:Ce scintillating calorimeter, reconstructs 3D RSP maps by a filtered-back projection algorithm. The imaging performances (i.e. spatial resolution, noise power spectrum and RSP accuracy) of the pCT system were assessed on a custom-made phantom, made of plastic materials with different densities ([0.66, 2.18] g/cm3). For comparison, the same phantom was acquired with a clinical xCT system. Main results: The spatial resolution analysis revealed the non-linearity of the imaging system, showing different imaging responses in air or water phantom background. Applying the Hann filter in the pCT reconstruction, it was possible to investigate the imaging potential of the system. Matching the spatial resolution value of the xCT (0.54 lp/mm) and acquiring both with the same dose level (11.6 mGy), the pCT appeared to be less noisy than xCT, with an RSP standard deviation of 0.0063. Concerning the RSP accuracy, the measured Mean Absolute Percentage Errors were (0.23+-0.09)% in air and (0.21+-0.07)% in water. Significance: The obtained performances confirm that the INFN pCT system provides a very accurate RSP estimation, appearing to be a feasible clinical tool for verification and correction of xCT calibration in proton treatment planning.File | Dimensione | Formato | |
---|---|---|---|
Fogazzi+et+al_2023_Phys._Med._Biol._10.1088_1361-6560_acd6d3.pdf
accesso aperto
Tipologia:
Post-print referato (Refereed author’s manuscript)
Licenza:
Creative commons
Dimensione
1.29 MB
Formato
Adobe PDF
|
1.29 MB | Adobe PDF | Visualizza/Apri |
Fogazzi_2023_Phys._Med._Biol._68_124001.pdf
accesso aperto
Tipologia:
Versione editoriale (Publisher’s layout)
Licenza:
Creative commons
Dimensione
1.44 MB
Formato
Adobe PDF
|
1.44 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione