The AG500 (Carstens Medizinelektronik GmbH) is widely used in electromagnetic articulography, as it allows the recording in 3D of the movements of the articulatory organs. However some anomalies in sensor position tracking were recognized and analyzed in previous works, which ultimately attributed them to numerical issues and not to external interferences. Recently a new upgraded model, the AG501, has been introduced. By means of both speech experiments and numerical measurements, we analyzed the reliability of the AG501 and compared it to that of the previous model, the AG500. While, in some scenarios, the positions obtained via the AG500 are afflicted by external perturbations, the AG501 showed a greater degree of precision in all the numerical experiments, and in the speech analyses that have been performed so far.

Electromagnetic articulography with AG500 and AG501 / Stella, M; Stella, A; Sigona, F; Bernardini, P; Grimaldi, M; Gili Fivela, B. - ELETTRONICO. - (2013), pp. 1316-1320. (Intervento presentato al convegno Interspeech 2013 tenutosi a Lyon nel 25-29 Agosto 2013).

Electromagnetic articulography with AG500 and AG501

Stella M
Primo
;
2013-01-01

Abstract

The AG500 (Carstens Medizinelektronik GmbH) is widely used in electromagnetic articulography, as it allows the recording in 3D of the movements of the articulatory organs. However some anomalies in sensor position tracking were recognized and analyzed in previous works, which ultimately attributed them to numerical issues and not to external interferences. Recently a new upgraded model, the AG501, has been introduced. By means of both speech experiments and numerical measurements, we analyzed the reliability of the AG501 and compared it to that of the previous model, the AG500. While, in some scenarios, the positions obtained via the AG500 are afflicted by external perturbations, the AG501 showed a greater degree of precision in all the numerical experiments, and in the speech analyses that have been performed so far.
2013
Proceedings of Interspeech
Lyon
Interspeech
Stella, M; Stella, A; Sigona, F; Bernardini, P; Grimaldi, M; Gili Fivela, B
Electromagnetic articulography with AG500 and AG501 / Stella, M; Stella, A; Sigona, F; Bernardini, P; Grimaldi, M; Gili Fivela, B. - ELETTRONICO. - (2013), pp. 1316-1320. (Intervento presentato al convegno Interspeech 2013 tenutosi a Lyon nel 25-29 Agosto 2013).
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11572/365667
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 14
  • ???jsp.display-item.citation.isi??? 15
social impact